Device evaluated by manufacturer: the device was returned for analysis.The main coil, the coil plunger and coil introducer were returned for this complaint.One section of the main coil was returned outside the coil introducer.The main coil returned was inspected and no damages were found.The coil plunger was advanced through the coil introducer without problems, no resistance was felt.The main coil was inspected and no damages were found.Dimensional inspection revealed that actual apex zap tip, base zap tip and primary coil (outer dimensional) were within ths specification.However, the fiber bundles were less than the specification.
|
Reportable based on device analysis completed on 19oct2019.It was reported that coil did not come out of the sheath.A 5mm/5.5mm vortx diamond - 18 was selected for use.During preparation, it was noted that the coil could not pass through the introducer sheath.The procedure was completed with another of the same device.No complications reported.However, device analysis revealed that the fiber bundles were incomplete.
|