The returned device was patent.The valve met the requirements for leak and reflux testing.The valve did not meet the requirements of pressure flow and preimplantation testing.Biological debris was observed on the interior of the valve.The instructions for use cautions, ¿shunt obstruction may occur in any of the components of the shunt system.The system may become internally occluded due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|