• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INTRAVIA CONTAINER EMPTY; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INTRAVIA CONTAINER EMPTY; CONTAINER, I.V. Back to Search Results
Catalog Number 2B8019
Device Problems Crack (1135); Disconnection (1171)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the ports of the unspecified quantity of 50ml intravia empty containers were cracked and falling out during spiking chemotherapy drugs.These issues were identified prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information : the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed and noted four (4) devices; however, no damage was observed on the photo samples.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA CONTAINER EMPTY
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9306278
MDR Text Key166398106
Report Number1416980-2019-06241
Device Sequence Number1
Product Code KPE
UDI-Device Identifier00085412003504
UDI-Public(01)00085412003504
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2B8019
Device Lot NumberDR19D02049
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/29/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/03/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-