Additional information: d10, g4, h3, h6 h3.Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found no defects.The reported condition was not confirmed.The investigation could not find any evidence of false negative detection.An additional, but unrelated failure mode was found when coil 4 showed transient failure.When the coil weld was manipulated, the coil resistance changed significantly.The mat was opened up and no evidence of fluid ingress was noted.The investigation could not determine the root cause of the customer's report.The additional failure of coil4 transient performance is due to the failure of the coil weld.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|