Ps323273.Section b1 report type and b2 amended due to further information received.Additional information was requested from the customer as part of the root cause analysis.The customer provided information regarding the patient's medical history relevant to this event.The patient's medical history included "seizures respiratory failure", "global developmental delay", and "polymicrogyria".The customer further reported that "patient remains on niv on a slow one way wean after the incident, however not due to the desaturation incident." results: during testing, the airvo 2 humidifier powered on and was found to be performing as expected.During the inspection, an incidental finding was that the audible alarm of the complaint airvo 2 humidifier was not working which is caused by a faulty speaker and electrical resistance testing has shown the speaker's resistance to being open circuit.A review of the airvo 2 humidifier device log did not present any evidence to suggest any issues with flow on the event date that would have triggered an alarm.Conclusion: we are unable to determine what has caused the reported event, as testing found the device to be functioning as expected and our review of the device log found no malfunction or flow issue that would have triggered an alarm.The pt101 airvo2 humidifier user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply".The airvo2 has visual and auditory alarms to warn operators about interruptions to patient's treatment.- prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system functionality check described in the alarms section.The alarm system functionality check instructs the user on how to test the alarm and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative." additionally, the user manual also warns the user: - the unit is not intended for life support.- appropriate patient monitoring must be used at all times.Loss of therapy will occur if power is lost.
|