The user facility returned the used device (including hf cable) to the service center for evaluation.The user facility also returned three brand new brand new devices.A visual inspection of the hf cable found no abnormalities on the appearance.The electrode loop has minor debris and was slightly bent as signs of use, no missing components noted.The distal insulation posts and shaft are in good condition.The used and new electrodes and the corresponding hf cables were checked together with a test working element esg-400 generator with the default settings saline cut 200watts and saline coag 120 watts.Both the used and new electrodes could cut and coag a tissue sample without any delayed firing.There was no intermittent problem observed.A review of the device history records was conducted as the manufacturing and quality control review was performed for the affected lot number without showing any non-conformities or deviations regarding the described issue.Based on the evaluation findings, the cause of reported event could not be determined as both the used and new devices tested and worked appropriately when tested with manufacturer test equipment.The device instruction manual indicates, to ensure only saline is used as the irrigant, and ensure both cable connections at the olympus resectoscope are secure.¿ if necessary, adjust the settings to the procedure and the required tissue effects.Increase the output power to a level that is needed for sufficient resection.If non-conductive irrigation fluids are used, there will be no cutting, coagulation or vaporization effect.¿ when using the resectoscope in combination with the electrosurgical generator plasmakinetic¿ superpulse or esg- 400, only use normal saline.
|
The manufacturer was informed that at the beginning of a transurethral resection of the prostate procedure, the customer was having a hard time firing the loop as the loop was sputtering and at times unresponsive.A second electrode loop was used but the same problem persisted.There was no breakage or sparking/arcing to the loops reported.Then tried to replace the generator but the same issue occurred.The generator settings were set to default 200watts (cut)-120watts (coag).The user reported there was no errors observed and that the generator did not malfunction as the user determined the problem was with the loop as once the user opened a different (medium) loop it worked and the procedure was completed.There was a 30-40 minute delay in the procedure.There was no serious injury or death reported.The electrode and connections to the working element were inspected prior to use, there were no anomalies noted.
|