MEDTRONIC EUROPE SARL COMPIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMC2QQ
Device Problem Pacing Problem (1439)
Patient Problems Arrhythmia (1721); Complete Heart Block (2627)
Event Date 10/22/2019
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced arrhythmia and heart block.The cardiac resynchronization therapy defibrillator (crt-d) was suspected of under pacing.The patient was observed on holter monitor with heart rates below the programmed lower pacing limit.However, threshold, sensing, and impedance tests appeared normal at follow up.Diagnostic testing/troubleshooting and reprogramming were performed.The device remains in use.It was also noted that the right ventricular (rv) lead exhibited suspected oversensing.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPIA MRI QUAD CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9324682
MDR Text Key168280109
Report Number9614453-2019-03863
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/14/2019
Device Model NumberDTMC2QQ
Device Catalogue NumberDTMC2QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076-52 LEAD, 6935M55 LEAD, 429878 LEAD
Patient Outcome(s) Required Intervention;
-
-