• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71940-01
Device Problem Low Readings (2460)
Patient Problems Hyperglycemia (1905); Vomiting (2144); Cramp(s) (2193)
Event Date 10/28/2019
Event Type  Injury  
Manufacturer Narrative
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported low scanned values while wearing the adc freestyle libre sensor compared to an adc meter.Sensor scans of 5mmol/l, 4mmo/l, /l, and 3.8mmol/l were compared blood glucose results of 18.7mmol/l, 13.3mmol/l, 14.2mmol/l, and 12.9mmol/l obtained on the adc meter.On (b)(6) 2019, the customer had contact with their hcp due to unspecified low scanned values and symptoms of cramping, vomiting, dry mouth, and "ketoacidosis symptoms." the hcp performed a blood glucose using hcp meter and obtained a "hi" reading (the customer was unable to provide the actual test results beyond "hi").The customer was treated with insulin (dose unknown) for a diagnosed with hyperglycemia.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
tara williamson
1360 south loop road
alameda, CA 94502-7001
5108644472
MDR Report Key9335677
MDR Text Key166800339
Report Number2954323-2019-08979
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71940-01
Was Device Available for Evaluation? No
Date Manufacturer Received10/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight67
-
-