Device identifier # (b)(4).The device was returned for evaluation.The device history record (dhr) was reviewed and no anomalies were observed.The device passed all required inspection points with no associated mrr's, variances or rework.With respect to the returned unit it has passed all specific functional testing requirements, when unit is properly positioned and put under pressure unit would not have slipped.Therefore, the reported complaint is not confirmed.The definite root cause of the reported complaint cannot be reliably determined.
|