The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome membrane leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.Kit lot h330 was reviewed.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h330 shows no trends.Trends were reviewed for complaint categories, pressure dome membrane leak, and alarm #18: system pressure.No trends were detected for these complaint categories.An investigation was performed based on the complaint description, the returned kit, and smart card.Review of the data on the returned smart card shows that an alarm #18: system pressure occurred.A leak from the system pressure dome was confirmed upon inspection of the received kit.The leak occurred because the diaphragm was partially unseated.When a pressure dome is not properly attached to its pressure sensor, because one of its latches is not secured in the circumferential groove, the diaphragm of the pressure dome will have room to move and become unseated.Evidence indicates the system pressure dome was most likely not properly installed during use.No further action is required at this time.This investigation is now complete.(b)(4).
|
The customer called to report that they experienced a pressure dome membrane leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer stated an alarm #18: system pressure occurred, and during troubleshooting the alarm, they observed blood leaking from the system pressure dome.The blood leak was observed after approximately 206 ml of whole blood had been processed.The ecp procedure was aborted, and the patient was reported to be in stable condition.The customer has returned the kit and smart card for investigation.
|