Model Number 9770086
Device Problem Failure to Align (2522)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed - the scanner is giving out inaccurate readings, the probe shows the vein once the nurse inserts the needle it's not where the probe shows.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the scanner is giving out inaccurate readings, the probe shows the vein once the nurse inserts the needle it's not where the probe shows was unconfirmed.The site rite prevue was visually inspected upon receipt and was found to be in good overall condition and functions normally.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed - the scanner is giving out inaccurate readings, the probe shows the vein once the nurse inserts the needle it's not where the probe shows.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE PREVUE PLUS
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key9359698
MDR Text Key167564503
Report Number3006260740-2019-03647
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741095450
UDI-Public(01)00801741095450
Combination Product (y/n)N
PMA/PMN Number
K120882
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770086
Device Catalogue Number9770086
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2019
Event Location Hospital
Date Manufacturer Received01/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-