Model Number N/A
Device Problem Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign report source: (b)(6).The device will not be returned for analysis, due to location of device is unknown; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: (b)(4).
 
Event Description
It was reported that the screw penetrated through the shell.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D4: udi#: (b)(4).The following sections were updated/corrected updated: b4, b5, d4, g4, g7, h2, h3, h6, h10.The event was confirmed with radiographs received.Review of the available records identified the following : the medial screw is fully seated within the bone and appears to have penetrated the acetabular cup.No other complications were noted.Dhr was reviewed and no discrepancies relevant to the reported event were found.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 20 MM LENGTH
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key9361839
MDR Text Key190845402
Report Number0002648920-2019-00855
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
PMA/PMN Number
K934765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00625006520
Device Lot Number64213153
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN SHELL.
-
-