• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Difficult to Insert (1316)
Patient Problem No Patient Involvement (2645)
Event Date 11/06/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that during freedom driver system check, there was difficulty when inserting the freedom onboard battery into the battery well.
 
Manufacturer Narrative
During investigation testing, the customer-reported issue was confirmed as there was difficulty inserting a freedom onboard battery into the right battery well.Despite this insertion difficulty, the freedom driver passed all functional testing criteria.The root cause of the reported issue was determined to be a malfunction of the right battery latch and the battery latch bars.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9362428
MDR Text Key167764688
Report Number3003761017-2019-00336
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2019
Date Manufacturer Received11/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-