ZIMMER BIOMET, INC. COMP PRIMARY STEM 13MM STD; PROSTHESIS SHOULDER Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Fracture, Arm (2351)
Event Date 08/27/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product's location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Product location is unknown.
 
Event Description
It was reported that bilateral patient underwent right reverse shoulder arthroplasty approximately 8 years ago.Subsequently, the patient underwent a revision procedure about 3 months ago due to fractured humeral tray.During the procedure they were unable to remove the stem and broke the patient's arm to get it out.
 
Manufacturer Narrative
(b)(4).Reported event was considered confirmed as the medical records stated that the trunnion from the humeral tray had broken.The stem was removed due to the trunnion fracture however is was well fixed and during removal the proximal humerus fractured.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
Upon reassessment of the reported event, this report should be voided as it was determined to be a duplicate (b)(4).Event will be captured within mfr number 0001825034-2022-00452.
 
Event Description
Upon reassessment of the reported event, this report should be voided as it was determined to be a duplicate (b)(4).Event will be captured within mfr number 0001825034-2022-00452.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP PRIMARY STEM 13MM STD
Type of Device
PROSTHESIS SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key9362775
MDR Text Key167607104
Report Number0001825034-2019-05286
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model NumberN/A
Device Catalogue Number113653
Device Lot Number399260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
Patient Weight82 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-