DEPUY ORTHOPAEDICS INC US PROSTALAC MOLD SZ3 240 RT; HIP INSTRUMENTS : MOLDS Back to Search Results
Catalog Number 254188000
Device Problems Component Missing (2306); Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/29/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> examination of the returned instrument confirmed the reported observation.The root cause is attributed to heavy usage and wear out.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that there was a missing metal post on prostalac mold, top mold moves when tightened.This was discovered during tray reassembly.No delay or patient harm occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSTALAC MOLD SZ3 240 RT
Type of Device
HIP INSTRUMENTS : MOLDS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6103142063
MDR Report Key9363418
MDR Text Key184368586
Report Number1818910-2019-117623
Device Sequence Number1
Product Code KWZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H000004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254188000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-