Model Number BI70000028230 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/14/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.It was determined the system would not boot up.The cmos battery was determined to be faulty.The cmos battery was replaced.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.There was no patient involvement.The imaging system would not get passed the initializing stage.
|
|
Manufacturer Narrative
|
Continuation of concomitant medical products: section 'suspect device' information references the main component of the system.Other relevant device(s) are: product id: bi71000165, serial/lot : na.If information is provided in the future, a supplemental report will be issued.
|
|