The manufacturer was informed that a patient reportedly expired and that the patient had creutzfeldt-jakob disease (cjd).The gi manager at the user facility further reported the scope had been used on other patients (not specified) but no other patient infections were reported from the use of this scope.The doctor performed a small bowel endoscopy with biopsy for mild esophagitis procedure on (b)(6) 2019.The patient's positive culture was detected on (b)(6) 2019.There was no scope malfunction reported.The scope was used with cold forceps.The scope was not cultured or sterilized since the date of the event.The scope was taken out of service at notification of patient's diagnosis, after being manual cleaned and high level disinfected.The gi manager also provided information regarding the reprocessing.Pre-cleaning is being performed immediately after a procedure as per manufacturer's pre-cleaning instructions.The scope is being brushed during manual cleaning.The scope is being leak tested.The effective concentration is being checked between each scope reprocessing.There are no issues with the aer.Flushing (air) into the channel of the scope is done after reprocessing.There have been no changes to the reprocessing personnel.The scope has been removed from service.
|