DEPUY MITEK LLC US OMNISPAN MENISCAL REPAIR 12DEG; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228141
Device Problems Premature Discharge of Battery (1057); Premature Activation (1484); Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/04/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/ or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the affiliate in (b)(6) that during a meniscus suture operation, it was observed that two plates on the omnispan meniscal repair 12deg device were deployed after the first firing.The procedure had to be completed using another device.No patient consequence and no surgical delay was reported.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: h6: method codes: device history lot
=
> a manufacturing record evaluation was performed for the finished device lot number (3l05593), and no non-conformances were identified.Investigation summary
=
> the complaint device is not being returned, it was implanted in the patient, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (3l05593), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNISPAN MENISCAL REPAIR 12DEG
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9370407
MDR Text Key207716661
Report Number1221934-2019-59756
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705010035
UDI-Public10886705010035
Combination Product (y/n)N
PMA/PMN Number
K150209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model Number228141
Device Catalogue Number228141
Device Lot Number3L05593
Was Device Available for Evaluation? No
Date Manufacturer Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-