• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXPRESS SD RENAL/BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 20188
Device Problems Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2019
Event Type  Injury  
Event Description
It was reported that stent dislodgement occurred.The target lesion was located in the right renal artery.A 5.0mmx19mmx90cm express sd renal/biliary drug-eluting stent was advanced through a tight stenosis.The stent would not advance through the tight stenosis.The device was being removed, it was noted during removal that the stent came off the balloon.The dislodged stent was snared out of the patient.The procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9371163
MDR Text Key167888166
Report Number2134265-2019-14427
Device Sequence Number1
Product Code NIN
UDI-Device Identifier08714729390480
UDI-Public08714729390480
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2022
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0023615185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-