Additional information: d10, g4, g7, h2, h3, h4, h6, h10.Device identifier # (b)(4).The device was returned for evaluation.The device history record (dhr) has been reviewed with no abnormalities related to the reported failure.The device passed all required inspection points with no associated mrr¿s, variances or rework.The reported complaint was not confirmed, no issues were observed.When unit is properly positioned and put under pressure unit would not have slipped.Unit has not had any service since purchased new in 2014.Pm maintenance and cleaning required at this time.The definite root cause cannot be reliably determined.Most probable root causes are: - incorrectly assembled device - improper technique used during procedure - inadequately maintained device used for procedure.
|