Catalog Number UNK HIP FEMORAL STEM
Device Problem Malposition of Device (2616)
Patient Problem No Code Available (3191)
Event Date 11/01/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient was being revised for repeated dislocation of their bipolar hip done by current surgeon's partner earlier this month.Previous cemeneted summit stem was found to be retroverted, however, surgeon elected to leave it rather than revise it due to patient age.Decision was made to convert to a total hip and constrained liner.No further patient information available.Doi: (b)(6) 2019; dor: (b)(6) 2019; left hip.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision depuy synthes of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL STEM
Type of Device
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9383365
MDR Text Key168228740
Report Number1818910-2019-118377
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL STEM
Was Device Available for Evaluation? No
Date Manufacturer Received12/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARTICUL/EZE BALL 28 +8.5 BL; UNKNOWN HIP FEMORAL STEM; ARTICUL/EZE BALL 28 +8.5 BL; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age89 YR
-
-