• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® DIGNISHIELD® STOOL MANAGEMENT SYSTEM; DIGNISHEALD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® DIGNISHIELD® STOOL MANAGEMENT SYSTEM; DIGNISHEALD Back to Search Results
Catalog Number SMS002
Device Problems Inability to Irrigate (1337); Obstruction of Flow (2423)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/04/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that water could not be injected into the flush port on the 3rd day of use.The device was replaced.
 
Event Description
It was reported that water could not be injected into the flush port on the 3rd day of use.The device was replaced.
 
Manufacturer Narrative
The device was not returned for evaluation.A potential failure mode could be ¿unable to flush catheter¿ with a potential root cause of ¿arm occluded or detaches¿.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿¿ catheter tube assembly (figure 1 includes collection bag) ¿ collection bag ¿ 60ml syringe ¿ lubricating jelly syringe (10ml) ¿ instructions for use ¿ 1 bottle (1oz) of medi-aire¿ biological odor eliminator ¿ tube clamp the bard® dignishield® sms catheter tube assembly consists of a catheter body and collection bag assembly that is primarily constructed of a proprietary copolymer material called permalene¿, bonded to a low-pressure retention cuff and trans-sphincteric zone (tsz) primarily constructed of silicone material.The permalene¿ catheter and collection bag material is designed to minimize permeation of gas and water vapor.The low-pressure retention cuff is designed to retain the device in the rectum.The tube opening at the cuff`along the drainage tube are three lumens, each with a separate access port.The green inflation port (¿inf(45ml)/inflate to 45ml¿) is used to inflate/deflate the cuff.The clear irrigation port (¿irrig¿) is used to infuse water at the end of the retention cuff and to provide access for the administration of medication.The purple flush port (¿flush¿) is used to infuse water through slits for the entire length of the drainage tube to assist drainage of fecal matter.(figure 2) a sample port on the drainage tube allows for the collection of stool samples through a slip-tip syringe.A piston valve connector located on the end of the drainage tube of the catheter attaches to the collection bag hub socket.When the collection bag is disengaged from the catheter, the catheter and bag automatically close to prevent spillage.A bag cap is provided to secure the contents of the collection bag when the catheter is removed.The 60 ml syringe and lubricating jelly syringe are used in the preparation and use of the catheter.The medi-aire¿ biological odor eliminator may be used as an air freshener in the room.Do not spray on patient or device.The tube clamp is used to retain medication during administration of medication.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® DIGNISHIELD® STOOL MANAGEMENT SYSTEM
Type of Device
DIGNISHEALD
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9386213
MDR Text Key177368819
Report Number1018233-2019-07608
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
PMA/PMN Number
K133251
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSMS002
Was Device Available for Evaluation? No
Date Manufacturer Received12/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-