• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 52 DISCOVERY RAD-8, HORIZONTAL; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 52 DISCOVERY RAD-8, HORIZONTAL; OXIMETER Back to Search Results
Model Number 22042
Device Problems No Audible Alarm (1019); Battery Problem (2885)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
Additional manufacturing narrative: other, other text: the returned device was evaluated.Visual inspection was performed and no external damage or defects were observed.During functional testing, the unit was found to shut down with no alarm triggered.A known good battery was installed and after 6 hours of monitoring function, the battery indicator started to blink and the low battery alarm was triggered.The original battery was charged for 4 hours and reinstalled into the device.The battery indicator illuminated 4 leds with a 100% charge.After 10 minutes of monitoring function, the battery indicator illuminated only 1 led and the unit shut down with no low battery alarm triggered.The failure was isolated to the unit's battery.A service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues related to this reported event.
 
Event Description
The customer reported that the device was not holding a charge and not giving a low battery alarm.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8, HORIZONTAL
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
stefan lissmann
52 discovery
irvine, CA 92618
9492977168
MDR Report Key9387429
MDR Text Key174877327
Report Number3011353843-2019-00127
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997005647
UDI-Public00843997005647
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number22042
Device Catalogue Number9190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-