Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d10, g4, g7, h1, h2, h3, h6, h10.The complaint is unconfirmed for one returned timberline articulating arm for the reported failure of functional failure: binding/jammed.Per the visual and functional investigation done, it was determined that the returned device does not make sounds when manipulated and the know is easy to tighten and loosen.No binding or jamming of the device was noticed as called out in the complaint, and the product is functionally conforming.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|