• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problems Failure to Form Staple (2579); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t93u4y.Date of event is 2019.Event day and event month were not provided.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 7 conforming clips.Upon testing, the jaws opened and closed without any difficulties.In addition, the device locked out as intended.The event/complaint described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device lot/batch number and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the device made an unusual noise when first operated and continued.One out of three staples worked.The clips were malformed.Replaced with new device and no issues.There was no delay to procedure and it is unknown how the procedure was completed.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
6107428552
MDR Report Key9398729
MDR Text Key197791072
Report Number3005075853-2019-23914
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEL5ML
Device Lot NumberT93U4Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2019
Date Manufacturer Received11/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-