MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JUGGERKNOT 1.0MM MINI 2-0 NDLS; SCREW, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JUGGERKNOT 1.0MM MINI 2-0 NDLS; SCREW, FIXATION Back to Search Results
Catalog Number 912076
Device Problem Failure to Eject (4010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that a juggerknot anchor did not deploy.No known adverse event was reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified juggerknot device was returned without the anchor.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERKNOT 1.0MM MINI 2-0 NDLS
Type of Device
SCREW, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9409393
MDR Text Key195424721
Report Number0001825034-2019-05422
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
PMA/PMN Number
K140908
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/24/2024
Device Catalogue Number912076
Device Lot Number324220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-