The product was discarded, therefore no product failure analysis can be conducted and device malfunction cannot be confirmed.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.A manufacturing record evaluation was performed for the finished device 30273608m number, and no internal action related to the complaint was found during the review.Manufacturer's ref.# (b)(4).
|
It was reported that a (b)(6) male patient underwent an atrial fibrillation ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and suffered st segment elevation and air embolism.At the beginning of the procedure, after the transseptal procedure, the physician noticed a drop in blood pressure of the patient.The physician checked for an effusion on intracardiac echocardiogram, but no effusion was present.A st-elevation myocardial infarction (stemi) was confirmed by checking the 12 lead electrocardiogram signals.The stemi seemed to resolve in about 10 minutes and the procedure was aborted.An echocardiogram was performed to look at the chamber function.Physician injected contrast through the arteries and confirmed that there was no longer any blockage.The patient was reported to be in stable condition.It is unknown if extended hospitalization was required.Physician¿s opinion on the cause of the event is that the thermocool® smart touch® sf bi-directional navigation catheter may not have been flushed before going into the body and some bubbles were introduced into the heart.During the event, biosense webster catheters were in the left atrium but no ablation had been performed yet.Per the thermocool® smart touch® sf bi-directional navigation catheter instructions for use (ifu) the catheter and tubing must be flushed per standard technique to ensure purging of trapped air bubbles and to verify that the irrigation holes are patent.
|