STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #3 R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5510-F-302
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Injury (2348)
Event Date 11/06/2019
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
It was reported that on (b)(6) 2019 a cemented cr femoral component was implanted without cement.No stryker rep was present for the procedure.Discovery was made (b)(6) 2019 when the rep noted an inventory discrepancy.Further investigation identified the case in which the component was used, and the fact that no cement was used.The surgeon was notified and the patient was revised on (b)(6) 2019.Rep provided the primary usage sheet and reported that no further information will be available.
 
Manufacturer Narrative
An event regarding revision due to user error involving a triathlon femoral component was reported.The event was confirmed based on provided implant sheet and label review.Method & results product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the most recent version of the ifu noted that "components labeled for ¿cemented use only¿ are to be implanted only with bone cement." a review of the label design for part 5510-f-302 was completed and noted that the following statement is referenced on the label: "warning: this device is intended for cemented use only".It was reported that the surgeon implanted this device without cement, which did not follow the instructions in ifu and product label.Thus, this event is determined to be caused due to user error.No further investigation for this event is required at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported that on (b)(6) 2019 a cemented cr femoral component was implanted without cement.No stryker rep was present for the procedure.Discovery was made (b)(6) 2019 when the rep noted an inventory discrepancy.Further investigation identified the case in which the component was used, and the fact that no cement was used.The surgeon was notified and the patient was revised on (b)(6) 2019.Rep provided the primary usage sheet and reported that no further information will be available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #3 R-CEM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9412274
MDR Text Key169210244
Report Number0002249697-2019-03941
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327040265
UDI-Public07613327040265
Combination Product (y/n)N
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model Number5510-F-302
Device Catalogue Number5510F302
Device Lot NumberDTJ7B
Was Device Available for Evaluation? No
Date Manufacturer Received01/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-