• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL AUGMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL AUGMENT Back to Search Results
Catalog Number UNK HIP FEMORAL AUGMENT
Device Problems Corroded (1131); Noise, Audible (3273)
Patient Problem No Code Available (3191)
Event Date 10/25/2009
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
"the literature article entitled, ""corrosion at the stem-sleeve interface of a modular titanium alloy femoral component as a reason for impaired disengagement"" written by christian r.Fraitzl, md, luis e.Moya, md, lorenzo castellani, md, timothy m.Wright, phd, and robert l.Buly, md published by the journal of arthroplasty vol.26 no.1 2011 accepted by publisher october 25, 2009 was reviewed.The article's purpose was to report on examination of 22 modular titanium alloy femoral components retrieved.0 to 8.8 years in vivo.All implants were the srom stem.The article does not discuss or provide any information regarding femoral head, acetabular cups or bearing surfaces.The main focus was the retrieved stem with examination of the sleeve/stem interface for corrosion.Each case is captured individually along with reason for revision and note of corrosion." this complaint captures specimen 12 from a (b)(6) with implantation period of 1.8 years 13 mm stem size and reasons for revision was malfunction/mechanical complication ("clicking" of the hip while sitting).Note of corrosion with black metal debris noted on stem and sleeve interface.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Visual examination of the provided photographs on (b)(4) confirmed the reported event.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL AUGMENT
Type of Device
HIP FEMORAL AUGMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 46581-0988
6107428552
MDR Report Key9412989
MDR Text Key184838364
Report Number1818910-2019-120369
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,literatur
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL AUGMENT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
-
-