• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Human-Device Interface Problem (2949); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a manufacturer representative regarding an implantable neurostimulator (ins) for gastrointestinal/pelvic floor and urinary dysfunction/sacral nerve stim.It was reported that the physician tried to turn the screw of the device and was unable to screw or unscrew it.A different battery was used in the procedure and the issue was resolved.There were no patient complications reported as a result of this event.On 2019-dec-06: no new information for the event description.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Analysis determined that the setscrew on the implantable neurostimulator (ins) was backed out beyond the point of being able to engage with the connector block.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9436883
MDR Text Key175449435
Report Number3004209178-2019-23331
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2019
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
Patient Weight75
-
-