The packaging of a optease retrieval 55cm catheter was found damaged after it was opened.The outer packaging was broken.The damage was noted upon initial receipt of the product.The product was not used.There was no reported patient injury.The device was stored as per the instruction for use.The product was returned for analysis.Per visual analysis of the picture received, a torn on the carton of the outer box package was observed on the front side of the unit.The connector of the stopcock valve was protruding through the outer box package of the unit.The two inner package pouches of the optease retr filter 55 were received for analysis folded inside a plastic bag.However, the outer box of the device was not returned for analysis.Per visual analysis of the returned pouches, no anomalies were found on the pouch package containing the retrievable vena cava filter set components.The pouch package containing the brite tip csi and vessel dilator components was received frayed/split/torn damaged.The connector of the stopcock valve was observed protruding through the film material of the pouch at the label level.No other anomalies were found.A product history record (phr) review of lot 17868949 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿packaging/pouch/box- frayed/split/torn - outer box¿ and ¿packaging/pouch/box- frayed/split/torn - inner package¿ were confirmed due to the confirmed protruding condition of the stopcock through one of the unit¿s pouches.However, the protruding condition observed could not be conclusively determined during the analysis.Shipping and /or handling factors may contribute to the failure as reported.According to the instructions for use, which are not intended as a mitigation of risk, ¿ remove the retrieval catheter from its packaging using sterile technique.Inspect the retrieval catheter upon removal from packaging to verify that it is undamaged.Warning: do not use a retrieval catheter that has been damaged in any way.If damage is detected, replace with an undamaged retrieval catheter¿.Neither the phr review analysis nor the product analysis results do not suggest that the observed stopcock protruding condition is related to the manufacturing process of the unit therefore, no corrective or preventative actions will be taken at this time, given that the reported and observed damages do not appear to be related to the manufacturing process.
|
As reported, the packaging of 55cm optease retrieval catheter was found damaged after it was opened.According to the image provided of the device, the outer box of the device appears to be torn.Per the manufacturer product evaluation of the returned device, the pouch package containing the brite tip csi and vessel dilator components was received with frayed/split/torn damaged.The connector of the stopcock valve was observed protruding through the film material of the pouch at the label level.There was no reported patient injury.The outer packaging was broken.The damage was noted upon initial receipt of the product.The product was not used.The device was stored as per the instruction for use.The device will be returned for evaluation.An image is available for review.
|