Catalog Number 311.34 |
Device Problems
Corroded (1131); Contamination /Decontamination Problem (2895)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that during the inspection of the loan kit, it was discovered that the kit was wet and there was a rust in three devices.There was no patient involvement reported.This report is for one (1) tap for 4.0mm cancellous bone screws.This is report 1 of 1 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Concomitant medical products: tap f/cancellousscr ø4 l110/85 (part # 311.340, lot # l643580, quantity 1).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H3, h4, h6: reviewing attached picture, the complaint description can be confirmed that on the article some rusty spots are visible.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot part: 311.340, lot: 9894189, manufacturing site: bettlach, release to warehouse date: 02.June 2016.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a review of the device history record.Device history lot part: 311.340, lot: 9894189, manufacturing site: bettlach, release to warehouse date: 02.June 2016.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H3, h6: investigation summary reviewing attached picture, the complaint description can be confirmed that on the article some rusty spots are visible.Product was not returned.Based on the information available, it has been determined that no corrective and preventive action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6 investigation summary.Visual inspection: the returned device was examined and minor wear and discoloration, drop-shaped spots, consistent with corrosion was noted on the device body.The observed damage is consistent with the reported complaint condition as such the complaint condition can be confirmed.All features related to the reported complaint condition were reviewed and no other issues were identified.Document/specification review: the manufacturing review shows that the production procedure was according to the specifications and there were no issues that would contribute to this complaint condition.Furthermore, as indicated in the manufacturing documents, the correct material (1.4112) was used according iso 7153-1.Summary: the received condition of the device is concordant with the reported complaint description and the complaint condition is therefore confirmed.However, since no fault could be found in material or during the production procedures.This failure is typically consistent with inadequate sterilization practices which allow residual moisture to remain on the device.Please note: regarding corrosion it can be stated, that all materials, including so called stainless steel as well, are conditionally only rust-resistant.The corrosion resistance is maintained only when the material is stored on a dry and metallic contactless condition.All metallic contacts on humid or wet condition create electrolytic reactions with contact corrosion as a result.Please find all reprocessing instructions for our products in the leaflet ¿important information¿.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.A conclusive determination could be reached and therefore the in the investigation flow listed remaining investigation steps are not required.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed as no product related issue could be detected device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|