MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAESTRO 4000 CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAESTRO 4000 CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 86248
Device Problem Device Remains Activated (1525)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/14/2019
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported that failure to halt ablation occurred.A maestro 4000 controller was selected for a atrial fibrillation procedure.Radio frequency (rf) failed to ceased upon release of foot pedal.Rf was ceased by manually overriding on the remote control by the attending cardiac scientist using the on/off button over three seconds after the pedal was released by the physician.The procedure was completed using the same device with no patient complications being reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO 4000 CONTROLLER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
STELLARTECH RESEARCH CORPORATION
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key9451221
MDR Text Key176242133
Report Number2134265-2019-15482
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number86248
Device Catalogue Number86248
Device Lot Number0103180713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2019
Initial Date FDA Received12/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/03/2018
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-