• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL TRUESPAN 24 DEGREE PEEK; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL TRUESPAN 24 DEGREE PEEK; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228152
Device Problem Disconnection (1171)
Patient Problem Not Applicable (3189)
Event Date 11/21/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Device was used for treatment, not diagnosis.Udi: (b)(4).
 
Event Description
It was reported by the affiliate via email that after the first anchor implementation with the truespan 24 degree peek the second anchor fell down.10 minutes delay reported.No patient consequences.Additional information provided by the affiliate reported the case was completed with another readily available truespan device and surgical intervention is not planned.It was also reported the device will not be returning for evaluation.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: h6: method codes: device history lot
=
> a manufacturing record evaluation was performed for the finished device lot number (2l16311), and no non-conformances were identified.Investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (2l16311), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 24 DEGREE PEEK
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9465740
MDR Text Key206361293
Report Number1221934-2019-59954
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026029
UDI-Public10886705026029
Combination Product (y/n)N
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Model Number228152
Device Catalogue Number228152
Device Lot Number2L16311
Was Device Available for Evaluation? No
Date Manufacturer Received01/28/2020
Patient Sequence Number1
-
-