The device was evaluated by the manufacturer, karl storz in (b)(4).As per the evaluation: the shaft was found to be bent, with a deflection of approximately 4 mm.This indicates high force application, which results in high leverage forces on the connection between the shaft and the tip.The broke off tip shows deformation, which indicates high force application.No indications for a material or manufacturing related issue were found during the investigation.The root cause most probably is too high force application, following high leverage forces on the connection of the tip of the shaft.A warning to not overload the instrument by exerting too much force is included in the ifu.The device was manufactured in may 2006.
|
As per a vigilance report filed with the (b)(6) in (b)(6), by our parent company in (b)(4): beveled end of cystoscope sheath became detached and remained in patient for 18 months.Patient had urological issues, pain, poor flow when passing urine.Patient returned to theatre with a what was thought to be a stone; when trying to break the stone apart, it was discovered to be the beveled end of the cystoscope sheath.
|