Model Number CI-1400-01 |
Device Problems
Intermittent Continuity (1121); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 11/22/2019 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issues did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual and photographic imaging inspections.System lock could not be obtained at any spacing.The device passed the electrical tests performed.The device failed the residual gas analysis test.This device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issue from one feedthru vendor.A corrective action was implemented.Feedthru assemblies from this vendor are no longer used.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|