Manufacturing site evaluation: the instrument arrived in a decontaminated condition.Surgery delay was longer than 15 minutes.The instrument arrived with broken off tip.The broken off tip was enclosed.Used test- and analysis- equipment: · microscope "keyence- vhx 5000 " eq.-nr.2000024840; · digital-camera "panasonic dmc tz8".At first we made a visual inspection of the dimensions of the fracture pieces and compared the values with the construction drawing.Length of the shaft (long piece): 228 mm.Length of the broken off piece: 29,3 mm.Amount: 257,3 mm.Target value: 257,4 mm.Therefore we received all pieces of the instrument.Because of the secondary damages, caused by the recovery, a investigation of the fracture surface of the tip makes no sense, so in the next step we investigated the fracture surface on the shaft.Here we found the typically signs of a multi axial stress condition of bending and torsion.Batch history review: the manufacturing documents have been checked and found to be according to specification valid during the time of production.There are no further complaints with this lot at hand.Conclusion and root cause: the root cause for the problem is most probably usage related.Rationale: the fracture surface exhibit the typically signs of a multi axial stress condition of bending and torsion.Without further knowledge about the circumstances, we suspect that the surgeon applied to high force to the instrument in conjunction with levering and bending.The instructions for use (ifu) points out, to avoid levering and bending during the threading.Corrective action: according to sop sa-de13-m-4-2-04-000-0 (corrective action & preventive action) a capa is not necessary.
|