C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE; BARD URETERAL STENT Back to Search Results
Model Number 777624
Device Problem Break (1069)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 11/23/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the bard urethral stent was found broken upon opening the package.
 
Event Description
It was reported that the bard urethral stent was found broken upon opening the package.
 
Manufacturer Narrative
The reported event was confirmed, the cause is unknown.The device failed to meet specifications.Evaluation report of the returned sample, submitted by futurematrix interventional, states that the bladder end (proximal) pigtail is broken from the stent.The pigtail straightener and the stent suture were not present with no additional observable deformities noted.The material of the stent at the break location showed no other sign of damage (stretching).The point of the break on the stent was located at the bladder end.The break was clean and the stent material was not stretched, which typically occurs when pulling on the stent.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿directions for use: 1.Determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage effi ciency and patient comfort.Submerge stent in sterile water to activate the coating.2.Insert the cystoscope then pass the guidewire* through the scope until the tip is in the renal pelvis.3.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.4.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement.5.Watch the distal end (bladder coil end) of the stent or the radiopaque, proximal end of the pusher.Stop advancing when the stent¿s distal end marker reaches the ureterovesical junction (uvj).**(see below for proper placement directions on the multi-length ureteral stent.) 6.Withdraw the guidewire slowly.The stent will form a pigtail automatically.7.Carefully remove the push catheter.*activate the guidewire coating according to the ¿instructions for use¿ found within the guidewire packaging.**multi-length ureteral stent placement: to accurately size this stent count the marker bands as it is being advanced into the ureter.The fi rst large band indicates the 22cm length.The second and third bands indicate 24cm and 26cm lengths respectively.The last large band is the 28cm length.If you need to place for the 30cm and 32cm lengths, use the attached suture or endoscopic forceps to gently pull back on the stent unwinding the coil from the kidney.Note: 1.Final adjustment, if necessary, can be made with endoscopic forceps.Stents can be removed easily by gentle withdrawal traction on the suture or by use of endoscopic forceps.2.Fluoroscopy facilitates stent placement; however, standard radiography may be used.3.The suture may be removed prior to placement or may be removed once indwelling by using an appropriate cystoscopic instrument.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Type of Device
BARD URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9476741
MDR Text Key178772247
Report Number1018233-2019-08011
Device Sequence Number1
Product Code FAD
UDI-Device Identifier10801741014540
UDI-Public(01)10801741014540
Combination Product (y/n)N
PMA/PMN Number
K983498
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/07/2023
Device Model Number777624
Device Catalogue Number777624
Device Lot NumberNGDQ3277
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2019
Date Manufacturer Received02/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-