ZIMMER BIOMET, INC. MINI COMP REV TRAY TRIAL; SHOULDER PROSTHESIS Back to Search Results
Catalog Number 110031431
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that trial humeral tray stem broke off during final explantation of trials.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified that the post of tray fractured and was not returned.Heavy gouges as shown on the device.Scanning electron microscope (sem) analysis shows fracture surface artifacts of hackle marks and river lines indicate an overload fracture.The score marks noted are likely post-fracture damage however this cannot be determined with certainty.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Upon reassessment of the reported event, it was determined this product should not have been reported under this mfr number.Additionally, this event is now considered to be not reportable.The initial report was forwarded in error and should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI COMP REV TRAY TRIAL
Type of Device
SHOULDER PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9481492
MDR Text Key189331047
Report Number0001825034-2019-05604
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K181611
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 06/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110031431
Device Lot Number64321423
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Was the Report Sent to FDA? No
Date Manufacturer Received06/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-