CLARET MEDICAL, INC. SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE Back to Search Results
Model Number CMS15-10C-US
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
Year of birth: (b)(6).
 
Event Description
It was reported that the distal filter was unable to be retrieved.Qualifying conditions: type one arch with a bifurcate right coronary artery.There is moderate tortuosity, mostly free from calcium, with some eccentric, ostial calcification of the innominate artery.Vascular access was obtained via the right radial artery.A transcatheter aortic valve replacement procedure was performed with the use of a sentinel cerebral protection system.During removal of the sentinel cerebral protection system, the #3 slider broke and the distal filter was unable to pull the distal filter into the system.The metal part slipped off the inner member.The physician pushed the handle together to disengage the left common carotid and the sentinel cerebral protection system was removed from the patient.No issues with the patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
CLARET MEDICAL, INC.
1745 copperhill parkway
santa rosa CA 95403
Manufacturer (Section G)
CLARET MANUFACTURING
1745 copperhill parkway
suite 1
santa rosa CA 95403
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9491117
MDR Text Key176635983
Report Number2134265-2019-15697
Device Sequence Number1
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public00863229000004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number19H01H20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/09/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient Weight94
-
-