STIMWAVE TECHNOLOGIES INC. FREEDOM SPINAL CORD STIMULATOR Back to Search Results
Model Number FR8A-RCV-A0
Device Problem Expulsion (2933)
Patient Problem Tissue Damage (2104)
Event Date 11/20/2019
Event Type  Injury  
Manufacturer Narrative
Immediately following notification, stimwave quality and the territory manager reviewed the events preceding the issue.The patient had a permanent procedure performed on (b)(6) 2019, in which one (1) freedom-8a receiver stimulator (fr8a-rcv-a0) and one (1) spare lead (fr8a-spr-b0) were implanted at the t8-t9 vertebral level.The territory manager confirmed that the implant procedures were performed in a sterile environment, sterile field handling protocols were used, the procedures were completed in accordance with the product instructions for use, and the sterile barriers of all product used were intact prior to implant.The procedures were completed without complications, and the territory manager maintained contact with the patient following implant.On an unknown date, the patient experienced a skin reaction at the implant site.The patient elected to visit a dermatologist and did not notify their implanting clinician.The dermatologist removed the surface of the skin that was affected.The patient was sent home the same day.On (b)(6) 2019, the patient attended a follow up visit with their implanting clinician and the territory manager.The territory manager noticed the mid-section of the patient's device was protruding through the skin where the patient's skin reaction had been prior.The implanting clinician did not observe any drainage or signs of infection.However, the decision was made to prophylactically explant the device that was exposed.The patient's second device remains implanted.The territory manager and the implanting clinician elected to turn off the device and wait for the patient's wound to heal before reprogramming the patient.No further complications were reported.The root cause of the issue cannot be traced to the device.The patient reported to be receiving therapy up until the moment the device was explanted.Device erosion is a known adverse event of spinal cord stimulators and the freedom scs system that is mitigated as far as possible in the product's risk management file.The source of the issue cannot be traced back to the device or procedure.The device did not fail to meet performance or safety specifications.Stimwave will continue to track and trend events.The root cause of the complaint is not attributed to device failure, the inability of the device to meet performance or safety specifications, or nonconformance to physical or functional device specifications.The stimwave product was not the source of the issue.The root cause of the issue is related to the treatment for the patient's skin reaction.The surface of the implant location was thinner due to the patient's treatment of the skin rection which subsequently lead to the patient's device erosion.Corrective action is not required to remedy the root cause of the complaint.The device did not fail to meet performance or safety specifications.Stimwave has confirmed that the issue is a known adverse event, mitigated as far as possible, and documented in the stimwave risk management file.Stimwave was in frequent contact with the territory manager from november 20, 2019, onward regarding the complaint and the root cause investigation.Stimwave confirmed that the product did not fail to meet performance and safety specifications.The source of the issue is attributed to patient's skin reaction.Stimwave has informed all parties that the product was not the source issue.In compliance with medical device reporting requirements and responsibilities, stimwave quality and its chief medical officer have determined that this issue is considered reportable as the event required medical or surgical intervention to prevent or preclude permanent impairment or damage.This event was reported to the united states food and drug administration (fda) on december 18, 2019.
 
Event Description
Stimwave quality has investigated the details regarding a complaint resulting from a device erosion reported to stimwave on november 20, 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM SPINAL CORD STIMULATOR
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
STIMWAVE TECHNOLOGIES INC.
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
STIMWAVE TECHNOLOGIES INC.
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
mary ann greenawalt
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key9492314
MDR Text Key186602631
Report Number3010676138-2019-00066
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182720
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberFR8A-RCV-A0
Was Device Available for Evaluation? No
Date Manufacturer Received11/20/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-