• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. CUP POSITIONER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. CUP POSITIONER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Naturally Worn (2988)
Patient Problem No Patient Involvement (2645)
Event Date 11/15/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign country: (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the threads on the tip of the inserter are stripped.No patient involvement or harm.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
Reported event was confirmed with product return.Visual inspection identified that the threaded tip was damaged and stripped.Damage was too severe for dimensional analysis.The strike plate exhibited strike marks consistent with use over a potential field age of approximately 4 years.No other damages were noted.As the reported event did not occur during a surgery, it is unknown if the adapter cap was used during previous procedures.Review of the device history records and receiving inspection records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP POSITIONER
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9494626
MDR Text Key190983260
Report Number0001822565-2019-05364
Device Sequence Number1
Product Code FZX
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00626001800
Device Lot Number62933040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-