Model Number 9770501
Device Problems Failure to Read Input Signal (1581); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has not been received by the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, the unit is freezing up in the ultrasound mode and the touchscreen is unresponsive.
 
Event Description
Per biomed, the unit is freezing up in the ultrasound mode and the touchscreen is unresponsive.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample(if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of this information the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of touchscreen is unresponsive was confirmed.The scanner was inspected the touchscreen is unresponsive, the cause of the issue is due to a front enclosure failure, leading the scanner to be unresponsive to user input.The device was serviced, tested and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key9497216
MDR Text Key176684933
Report Number3006260740-2019-03954
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
PMA/PMN Number
K15254
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2020
Event Location Hospital
Date Manufacturer Received04/13/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-