SMITH & NEPHEW, INC. RETROGRADE DRL 10MM; BIT, SURGICAL Back to Search Results
Catalog Number 72204046
Device Problem Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 11/27/2019
Event Type  malfunction  
Manufacturer Narrative
Foreign zip code: (b)(6).
 
Event Description
It was reported that during an unknown procedure, the acufex trunav did not close after flipping inside and finally broke.The procedure was completed with a backup device with no significant delay or patient injury reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H6: the reported 10mm retrograde drill, intended for use in treatment, has not been returned for evaluation.Without the reported product a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the cutting blade would not retract.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: not advancing the tip of the drill to a location that allows retraction of the cutting blade after use.Bone lodged within the drill housing.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the manufacturing and complaint records was performed, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 10MM
Type of Device
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key9497283
MDR Text Key174180350
Report Number1219602-2019-01589
Device Sequence Number1
Product Code GFG
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/05/2024
Device Catalogue Number72204046
Device Lot Number4888953
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-