SMITHS MEDICAL ASD, INC. LEVEL 1 H-1200 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
|
Back to Search Results |
|
Device Problems
Device Alarm System (1012); False Alarm (1013)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/19/2019 |
Event Type
malfunction
|
Event Description
|
Information was received that a new smiths medical level 1 h-1200 fast flow fluid warmer was "in-serviced" by the account manager at the hospital.It was reported that three alarms (a disposable, recirculating solution, and an over temp) were displayed on the device.The reporter also stated the disposable was placed correctly, the recirculating solution was full, and the temperature of the unit was only at 27c.No adverse effects were reported.
|
|
Manufacturer Narrative
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product sample was received; therefore, visual and functional testing could not be performed.A review of the photos provided confirmed the unit displayed the error light; however, the reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.
|
|