SMITHS MEDICAL ASD, INC. LEVEL 1 H-1200 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Device Problems Device Alarm System (1012); False Alarm (1013)
Patient Problem No Patient Involvement (2645)
Event Date 11/19/2019
Event Type  malfunction  
Event Description
Information was received that a new smiths medical level 1 h-1200 fast flow fluid warmer was "in-serviced" by the account manager at the hospital.It was reported that three alarms (a disposable, recirculating solution, and an over temp) were displayed on the device.The reporter also stated the disposable was placed correctly, the recirculating solution was full, and the temperature of the unit was only at 27c.No adverse effects were reported.
 
Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product sample was received; therefore, visual and functional testing could not be performed.A review of the photos provided confirmed the unit displayed the error light; however, the reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 H-1200 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
3350 granada avenue north
suite 100
oakdale MN 55128
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key9497374
MDR Text Key177563846
Report Number3012307300-2019-07288
Device Sequence Number1
Product Code BSB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
LEVEL 1 TRAUMA FAST FLOW PRESSURE CUFF.
-
-