Model Number DL-330-10-K
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted lead was not returned to neuropace for analysis.However, review of the device ecog and lead impedance confirmed that the ecog data are suggestive of a potential lead break.
 
Event Description
On (b)(6) 2019, intermittent artifact was identified through ecog review, suggestive of a possible lead break in the left depth lead connected to port 2.Further review of the lead impedance and ecog data was performed and was consistent with a lead break.During a routine rns neurostimulator replacement procedure, the lead was explanted and a new lead was implanted and connected to port 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key9501373
MDR Text Key196218459
Report Number3004426659-2019-00062
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005045
UDI-Public010085554700504517210606
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDL-330-10-K
Device Catalogue Number1007605
Device Lot Number26124-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age21 YR
-
-