TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 816571 |
Device Problem
Fracture (1260)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, when the roller pump was opened, the gut assembly rod was found to be broken along with the screw.
|
|
Event Description
|
Upon receipt of the device, the user reported that the roller pump was making a metallic sound with vibration while running.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed by information obtained from the manufacturer's subsidiary.No parts were returned for further evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Updated blocks: d9, h3, and h6.The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed that the threaded portion of the roller pump gut assembly shaft broke off at the nut and washer.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|