Model Number N/A
Device Problem Gas/Air Leak (2946)
Patient Problem No Patient Involvement (2645)
Event Date 11/26/2019
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformances related to the reported event were noted.A getinge service representative reported that the issue was fixed by exchanging the special seal.The iabp is now ready for clinical service, but has not yet been delivered to the customer.No further investigation is required.(b)(6).
 
Event Description
It was reported that during an ongoing repair that was being performed by a getinge service representative under mfg report number 2249723-2019-01548, the cardiosave intra-aortic balloon pump (iabp) had a helium gas leak that exceeded the criteria value of 20 psi in 5 minutes.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID INT TYP B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key9504320
MDR Text Key191562094
Report Number2249723-2019-02070
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-65
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-