Product event summary: analysis was unable to confirm the customer comment that the "pacer turned." however, analysis did reveal that errors had occurred, the main printed circuit board (pcb) was out of specification, the lower case was broken, the main seal was c ontaminated, the displayed was "bleeding" (out of specification electrical), the display wires were pinched but wire insulation was not compromised, and three case screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device then passed all functional tests.If information is provided in the future, a supplemental report will be issued.
|