ST. JUDE MEDICAL, INC. VIEWFLEX¿ XTRA ICE CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR
|
Back to Search Results |
|
Model Number D087031 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
Vascular Dissection (3160)
|
Event Date 12/04/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results, method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During a procedure a dissection of the inferior vena cava (ivc) occurred.While inserting the ice catheter, resistance was noted and the patient was diagnosed with the dissection of the ivc.A balloon was inflated at the site and the procedure was postponed for a later date.The patient remained in stable condition.There were no performance issues with any abbott device.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported dissection could not be conclusively determined.
|
|